Spirometry is one of the most common tests in medicine and one of the most often misread. It is also a strong fit for AI: the inputs are structured, the standards are written down and the volume is high. That makes it easy to ship a model and hard to know whether it is right. This guide is for health-system clinical AI leads, device teams and model builders who need a defensible answer.

Why does spirometry interpretation need AI at all?

Because humans disagree more than most people assume. In a 2019 study in the European Respiratory Journal, 120 pulmonologists from 16 European hospitals each interpreted the same 50 pulmonary function test cases, for 6,000 interpretations in total. They matched the ATS/ERS pattern in 74.4% of cases and reached the correct diagnosis in 44.6%. The AI software in the study matched the pattern in every case and was correct on diagnosis in 82%.

Agreement between the pulmonologists was moderate for pattern (kappa 0.67) and low for diagnosis (kappa 0.35). That last number is the one evaluators should remember. If specialists disagree that much, a single clinician's read is a weak reference standard, and a test set built that way will mislead you.

What are the two jobs an AI spirometry tool does?

Most products do one or both of these. Evaluate them separately.

JobWhat the AI doesReference standardWhat to measure
Quality gradingJudges whether each maneuver is acceptable and repeatable, and grades the sessionRespiratory therapist and pulmonologist review against ATS/ERS acceptability and repeatability criteriaAgreement on accept/reject per maneuver, session grade agreement, false passes
InterpretationClassifies the pattern (normal, obstructive, restrictive, mixed, non-specific) and suggests likely diagnosesAdjudicated pulmonologist consensus with clinical context and, where available, full PFTs and imagingPattern accuracy, diagnosis accuracy, severity agreement, referral safety

Quality grading matters more than it looks. A model that interprets a bad maneuver confidently produces a clean-looking wrong answer. In primary care, where many tests are run by staff who do spirometry occasionally, quality is often the weak point.

Which standards should the AI follow?

Two documents anchor most evaluations. The ATS/ERS standardization of spirometry sets acceptability and repeatability criteria for each maneuver. The ERS/ATS technical standard on interpretive strategies, published in the European Respiratory Journal in 2022, updated the 2005 interpretation standards. It covers reference equations, limits of normal, bronchodilator responsiveness, severity and the classification of common impairments, and it says results should be read with clinical context and measurement uncertainty in mind.

Ask the vendor which version of each standard the model follows, and which reference equations it uses. A switch in reference equations can move patients across the limit of normal without any change in their lungs. If your health system changes equations, treat it as a model change and re-evaluate.

How should you build the test set?

We build spirometry test sets around these rules.

  1. Use real tests from your own sites and devices. Flow-volume loops vary by spirometer and by technician.
  2. Include poor-quality sessions on purpose. The model must reject them, not interpret them.
  3. Cover the full case mix. Normal, obstructive, restrictive patterns, mixed and non-specific results, borderline values near the limit of normal and post-bronchodilator tests.
  4. Stratify by age, sex and height range. Reference equations behave differently at the extremes, and so do models.
  5. Keep it independent. No test that was used to train or tune the model.

For sampling detail, see our guide to representative AI test data and how to build a golden dataset for medical AI.

Who should grade the reference standard?

Respiratory therapists and pulmonary function technologists grade maneuver quality. They run these tests every day and know a cough artifact or early termination on sight. Pulmonologists grade interpretation, with at least two independent readers per case and a third to adjudicate. Where full PFTs, imaging or a confirmed diagnosis exist, give them to the adjudicator, not the first-round readers.

Measure grader agreement before you score the AI, the same way the 2019 study did. Our piece on inter-rater reliability in medical AI explains how.

What metrics should a clinical AI lead ask for?

Use this checklist when reviewing a vendor report or running your own study.

  • Per-maneuver agreement on acceptability, and the false-pass rate on bad maneuvers
  • Pattern accuracy against adjudicated consensus, with confidence intervals
  • Diagnosis accuracy by condition, not one blended number
  • Severity agreement for obstruction and restriction
  • Missed referrals: abnormal results the model called normal
  • Performance by age band, sex and site
  • Behavior near the limit of normal, where small errors change the label
  • Clinician workload: how often users override the AI, and why

Check for bias explicitly. Lung function reference standards have changed in recent years, and performance gaps across patient groups can hide inside a strong average. Our guide on how to test clinical AI for bias covers the method.

How do you keep evaluating after go-live?

Spirometry AI drifts for ordinary reasons. Sites buy new spirometers. Software vendors update models. Health systems change reference equations or reporting templates. Each of these can shift results without anyone noticing.

Run a quarterly blinded audit: sample AI-read sessions, have them graded by respiratory therapists and pulmonologists and compare against your go-live baseline. Re-run the full test set after any device, software or equation change.

Where do model teams building spirometry AI go wrong?

Three mistakes come up again and again. Teams train and test on curated hospital lab data, then deploy into primary care, where maneuvers are shorter, noisier and run by occasional users. They label with a single reader, which bakes one clinician's habits into the model. And they treat reference equations as fixed, so the model learns a limit of normal that the health system later changes. Each mistake is cheap to avoid at the data stage and expensive to find after launch.

How Acumen helps

We build evaluation data for respiratory and lung function AI with credential-verified pulmonologists and respiratory therapists. That includes quality-graded maneuver sets, adjudicated interpretation labels, bias-stratified test sets and recurring audit sets for post-launch monitoring. Every label records the reviewer's specialty and how disagreements were resolved. See our medical model evaluation service for how an engagement runs, or validate an existing dataset you already hold.

Request a sample dataset

For related reading, see our guides to evaluating AI lung nodule detection, inter-rater reliability in medical AI and building a golden dataset for medical AI.

FAQ

Is AI better than pulmonologists at reading pulmonary function tests?

In one 2019 European Respiratory Journal study, AI software matched the ATS/ERS pattern in every case and reached the right diagnosis in 82%, against 44.6% for 120 pulmonologists. That is one study on 50 cases. Results on your own patients, devices and reference equations need local testing before you rely on them.

What does AI spirometry software actually do?

Most tools do two things. They grade whether each maneuver meets ATS/ERS acceptability and repeatability criteria, and they interpret the result by classifying the pattern and suggesting likely diagnoses. Some also draft the report. Evaluate quality grading and interpretation separately, because a model can be strong at one and weak at the other.

How do you build a test set for AI spirometry interpretation?

Use real sessions from your own sites and spirometers. Include poor-quality tests, the full range of patterns, borderline values near the limit of normal and post-bronchodilator results. Stratify by age, sex and height range, and keep every test independent of the model's training and tuning data.

Who should grade spirometry for an AI evaluation?

Respiratory therapists or pulmonary function technologists grade maneuver quality. At least two pulmonologists independently interpret each case, with a third adjudicating disagreements using full PFTs or imaging where available. Measure agreement between graders first, so you know how strict a standard the AI can fairly be held to.

Can AI spirometry help in primary care?

It can, mainly by catching poor-quality maneuvers and giving occasional users a consistent interpretation. The risk is false reassurance on borderline or badly performed tests. Before rollout, test the tool on sessions run by your own primary care staff and track missed abnormal results as the main safety metric.

How often should AI spirometry be re-evaluated?

Audit a blinded sample of AI-read sessions at least quarterly. Re-run the full test set after any new spirometer, software update or change in reference equations, because each can shift labels near the limit of normal. Compare every audit against your go-live baseline and investigate drops by site and patient group.