How to evaluate AI in radiation oncology
Acumen Editorial · Updated November 27, 2026
Auto-contouring is the most mature AI use in radiation oncology. It is also the easiest to over-trust. A contour can score well on overlap and still move dose to the wrong place. Another can look wrong and change nothing that matters.
The field now has real deployment data. At the SIIM annual meeting in June 2026, Penn's director of AI, Rafe McBeth, PhD, described auto-contouring across a 15-site network, per AuntMinnie. Clinical use reached 95%. Dice scores ranged from 0.80 to 0.85 across hundreds of structures in more than 4,000 patients. The tool returned about 15,000 clinical hours over three years.
Vendors are also shipping faster. GE HealthCare's MIM Contour ProtégéAI+ 2.0 received FDA 510(k) clearance with a predetermined change control plan, reported June 4, 2026. A PCCP lets the vendor add regions and modalities without a new clearance each time. Your evaluation has to keep up.
Why isn't a Dice score enough?
Dice measures overlap between two shapes. It does not tell you whether a difference sits next to the spinal cord or in harmless tissue. It also depends on the reference. If one physician drew the reference, you are measuring agreement with that physician, not with the truth.
A Cleveland Clinic study presented at ASTRO 2025 shows the gap. Researchers compared two AI systems with physician contours on post-implant CT in 100 prostate brachytherapy patients. One system's median prostate volume was more than double the physicians'. Yet the contours showed no significant difference in predicting clinical outcomes, per Cleveland Clinic's Consult QD. Geometry and clinical meaning can diverge in both directions.
Use geometric metrics as a screen. Then measure what clinicians do and what dose does.
What should you measure?
Build your evaluation around four questions.
How often do clinicians edit, and how much? Record, per structure, whether the oncologist accepted the contour, made minor edits or redrew it. Time each edit. This is the number that drives workflow value and the one your clinicians will believe.
Do the edits change the plan? Re-plan with the AI contour and with the edited contour. Compare dose to targets and organs at risk against your clinical constraints. A large geometric edit with no dose impact is a lower risk than a small one that pushes a constraint over its limit.
Does performance hold across sites and patients? Validate each disease site and imaging modality on its own. Include post-surgical anatomy, implants, contrast and unusual patient positioning. Penn noted that vendor roadmaps did not fully match its priorities for tumor and target contouring, so test the structures you use, not the ones in the brochure.
Does it stay good after updates? Re-run a fixed test set after every model version, scanner change or protocol change. With PCCP-enabled products, versions can change more often than your last commissioning report assumed.
Who should grade the outputs?
Match the grader to the decision.
- Radiation oncologists by disease site grade targets and organs at risk. A head and neck specialist should not be the only reviewer for pelvic structures.
- Medical physicists run dosimetric comparisons, commissioning and ongoing QA.
- Dosimetrists report how edits affect planning time and plan quality.
Use at least two independent graders on a sample of each structure. Report how often they agree. Without that, you cannot tell model error from normal variation between experts.
Evaluation checklist for radiation oncology AI
| Check | What to record | Who owns it |
|---|---|---|
| Reference set | Multi-reader contours per structure, with reader credentials | Radiation oncologists |
| Geometric screen | Dice, Hausdorff distance and surface distance per structure | Medical physics |
| Clinical acceptability | Accept, minor edit or redraw per structure, with edit time | Radiation oncologists |
| Dosimetric impact | Target coverage and OAR constraint changes, AI vs. edited contour | Medical physicists |
| Subgroups | Results by site, modality, post-op anatomy and implants | Physics and oncology |
| Inter-rater agreement | Agreement between independent graders on a shared sample | Evaluation lead |
| Update control | Fixed test set re-run after each model or protocol change | Medical physics |
| Ongoing QA | Automated flags plus periodic human review | Physics and QA |
Automated QA can help with the last row. A 2024 arXiv study used the AAPM Thoracic Auto-Segmentation Challenge dataset, covering lungs, heart, esophagus and spinal cord, to train a one-class classifier that flags low-quality AI contours. Tools like this triage review. They do not replace a graded reference set.
How should you run a pilot?
Start with one disease site and a fixed sample of recent cases. Have the AI contour every case. Have two oncologists grade each structure blind to each other, and have a physicist re-plan the cases with meaningful edits.
Report three numbers to your committee: share of structures accepted without edits, median edit time per case and the number of plans where edits changed a dose constraint. Then move to the next site.
Plan for the downstream effect. Penn reported that faster contouring saved about 30 minutes per patient for organs at risk, but the gain plateaued. Patients arrived faster and the bottleneck moved down the workflow. Measure the whole path from simulation to treatment, not only contouring time.
How do you set acceptance criteria?
Write the criteria before you see results. Teams that set thresholds after the pilot tend to move them to fit the data.
Set criteria per structure class, not one number for the whole product. Organs at risk near a tight constraint, such as the spinal cord or brainstem, deserve a stricter bar than structures with wide dose margins. Targets deserve the strictest bar of all, because a missed margin affects tumor control.
A practical set has three parts:
- A geometric floor. A minimum overlap and maximum surface distance per structure, used to flag cases for closer review rather than to approve them.
- A clinical bar. A minimum share of structures accepted without edits or with minor edits, judged by oncologists for that site.
- A dose rule. Zero cases where an unedited AI contour would have let a plan pass a constraint it should have failed.
Write down what happens when a criterion fails. The usual answer is to keep the tool for that site with mandatory review, or to pause it until the vendor's next version passes the same test set.
Keep the criteria, the test set and every grading record together. That bundle is what you show a quality committee, and what you re-run after an update.
What should you ask an auto-contouring vendor?
- Which structures, sites and modalities were validated, and on how many institutions' data?
- Who drew the reference contours, with what credentials, and how many readers per structure?
- What were edit rates and dosimetric impact, not only Dice?
- How do you handle post-surgical anatomy, implants and contrast?
- What changes can your PCCP cover, and how will we be told before a model changes?
- Will you support a local validation on our cases before go-live?
How Acumen helps
Acumen supplies medical model evaluation from verified clinicians matched to the task. For radiation oncology AI, we build multi-reader reference sets with radiation oncologists by disease site and medical physicists, grade AI contours on written rubrics, record edits and rationale per structure and report agreement on every batch. You get a fixed test set you can re-run after each model update. For the treatment decision side of cancer care, read how to evaluate oncology AI decision support.
FAQ
What is AI in radiation oncology used for today?
The most common use is auto-contouring: drawing organs at risk and, increasingly, targets on CT and MR images for treatment planning. Other uses include automated planning, plan quality checks and adaptive radiotherapy. Penn reported 95% clinical use of auto-contouring across a 15-site network, which shows how routine the contouring step has become in some centers.
Is a Dice score enough to validate auto-contouring?
No. Dice measures shape overlap against one reference, so it misses where the difference sits and whether it changes dose. A Cleveland Clinic study found AI prostate volumes far larger than physicians' with no significant difference in outcome prediction. Pair geometric metrics with oncologist edit rates and dosimetric comparisons on real plans.
Who should grade AI contours?
Radiation oncologists who treat that disease site should grade targets and organs at risk. Medical physicists should run dosimetric comparisons and commissioning. Dosimetrists can report effects on planning time and plan quality. Use at least two independent graders on a sample of each structure and report their agreement, so expert variation is not mistaken for model error.
How do we measure dosimetric impact of AI contours?
Plan each case twice: once with the AI contour and once with the oncologist-edited contour. Compare target coverage and dose to organs at risk against your clinical constraints. Count the cases where edits changed whether a constraint was met. That number tells you where AI errors matter clinically and where review effort should focus.
What does a PCCP mean for radiation oncology AI evaluation?
A predetermined change control plan lets a vendor update a cleared device within agreed limits without a new FDA submission for each change. For buyers, models may change more often. Keep a fixed local test set, ask vendors to notify you before updates and re-run the test after each version change.
How long should a radiation oncology AI pilot take?
Scope it by site rather than by calendar. Start with one disease site and a fixed sample of recent cases, graded by two oncologists and checked dosimetrically by a physicist. Report acceptance rate, edit time and constraint changes. Expand site by site once each passes your acceptance criteria.