Hematologists and hematopathologists reviewing digital blood smear and marrow AI evaluations

Healthcare · · 8 min read

AI in hematology: how to evaluate blood smear, bone marrow and decision support tools

Acumen Editorial · Updated November 26, 2026

To evaluate AI in hematology, test it against slides and cases whose ground truth was set by several hematologists or hematopathologists, not one reader. Report accuracy by cell class, not only overall. Weight blasts, atypical lymphocytes and other rare cells heavily, and confirm the tool flags what it cannot classify for manual review.

What does AI in hematology actually do today?

Most hematology AI in clinical use sits in the morphology lab. Digital morphology analyzers scan a stained slide, locate cells and pre-classify them, and a technologist or pathologist confirms or corrects each call. The AI drafts; the human signs.

Scopio Labs' Full Field Peripheral Blood Smear application is a useful reference point. HCPLive reported its FDA clearance for in vitro hematology diagnosis in 2020. The tool pre-classifies 200 white blood cells, gives a platelet pre-estimate and supports red cell morphology review. The company later received FDA de novo clearance for a bone marrow aspirate application that, according to Fierce Biotech, highlights cell types and helps estimate blast and plasma cell counts to "assist trained pathologists."

A second layer is arriving on top: language models that summarize a smear, a flow report and a CBC trend into a differential, or suggest a next test. That layer needs a different kind of evaluation, and we cover it below.

How accurate is AI blood smear analysis?

It depends on the cell class, and averages hide the cases that matter. A 2025 study in Clinical Chemistry and Laboratory Medicine compared two digital analyzers on 461 routine slides. One reached 90.0% overall pre-classification accuracy; the other reached 45.5%. Both did worse when abnormal cells were present.

The blast results are the part every buyer should read. One analyzer over-called blasts, with a 30.5% false positive rate. The other missed them, with a 42.1% false negative rate. Both had high false negative rates for atypical lymphocytes. The authors recommend manual inspection when blasts are observed, in line with International Council for Standardization in Haematology guidance.

So a single accuracy figure is not an answer. A tool can be excellent on segmented neutrophils and still miss the one cell that changes a diagnosis.

What should a hematology AI test set include?

Build the test set around the decisions the lab makes, then fill it with the hard cases routine workload rarely supplies.

Test set layerWhat it containsWhy it matters
RoutineNormal and reactive smears in the lab's real mixMeasures day-to-day pre-classification and review time
Enriched abnormalBlasts, promyelocytes, atypical and reactive lymphocytes, plasma cellsExposes false negatives that drive missed diagnoses
Morphology trapsSmudge cells, nucleated red cells, giant platelets, clumps, poor stainsTests whether the tool flags rather than guesses
Bone marrowAspirates with known blast and plasma cell percentagesChecks counts near diagnostic thresholds
Pre-analytic variationDifferent stainers, slide makers, sites and staining agesShows whether performance holds outside the vendor's lab
Case-level vignettesSmear plus CBC, flow and history for LLM decision supportGrades the reasoning, not just the cell call

Keep the test slides out of every training and tuning set. FDA's January 2025 draft guidance on AI-enabled device software asks sponsors to show test-data independence, and labs should hold vendors to the same standard. Our guide to representative AI test data covers how to size and stratify the set.

Who should set the ground truth?

Ground truth in morphology is a consensus, not a fact read off a slide. Experienced reviewers disagree on bands versus segmented neutrophils, on reactive versus atypical lymphocytes and on borderline blast counts. If one person labels the slides, the tool is graded against that person's habits.

Use at least two independent reviewers per cell or case, with a third hematopathologist adjudicating disagreements. Record agreement statistics before you score the AI, so you know the ceiling. Our explainer on inter-rater reliability in medical AI shows which statistics to report and how to read them. Freeze the adjudicated set as a golden dataset and version it.

How do you evaluate hematology decision support built on LLMs?

Cell classification has a right answer per cell. Decision support does not. When a model reads a smear summary, a CBC and a flow report and proposes "consider acute promyelocytic leukemia, start coagulation workup now," the question is whether a hematologist would act on that reasoning.

Grade these outputs with case-specific rubrics written by hematologists, the approach OpenAI's HealthBench made standard: physician-written criteria per case rather than one generic scale. Score at least these:

  • Urgency recognition. Does it escalate findings that need same-day action, such as suspected APL or schistocytes with thrombocytopenia?
  • Differential quality. Is the differential ordered sensibly for this patient, with the dangerous options present?
  • Next test. Is the recommended test the one a hematologist would order next, and not five tests at once?
  • Uncertainty. Does it say when morphology alone cannot decide and a marrow or flow study is needed?
  • Scope. Does it avoid treatment dosing it was never validated for?

Report worst-case performance alongside averages. A tool that handles 95% of routine anemias but misses a leukemic picture is not ready.

How does AI change the review workflow in the hematology lab?

Pre-classification changes what the reviewer does. Instead of finding and naming each cell, the technologist confirms a gallery the software has already sorted. That is faster, and it carries a known risk: people accept what is already on the screen. A blast filed under "lymphocyte" in a tidy gallery is easier to miss than a blast found under the microscope.

Test for that directly. Run a reader study where the same reviewers grade the same slides twice, once with the AI's pre-classification and once without, separated by a washout period. Compare error rates, not just time. If reviewers miss more abnormal cells with the AI than without it, the tool has made the lab faster and worse.

Remote review adds a second question. Digital slides let a hematopathologist review a smear from another site, which helps small labs. It also means the reviewer may never see the glass. Confirm that image quality at the remote screen supports the calls the lab is asking them to make, especially red cell inclusions and platelet morphology.

Finally, measure what happens to the cases the tool flags as unclassified. A flag rate the lab cannot staff becomes a queue nobody clears, and the safety net disappears.

What acceptance criteria should a lab set before go-live?

Set the criteria before testing, and write them down. A practical starting checklist:

  1. Per-class sensitivity and positive predictive value for every cell class the tool reports, with confidence intervals.
  2. A minimum blast sensitivity agreed with the hematopathologists, and a rule that any blast call routes to manual review.
  3. A flag rate for "unclassified" cells that the lab can staff.
  4. Agreement with adjudicated marrow counts near diagnostic cut-offs.
  5. Performance by site, stainer and slide maker, not pooled.
  6. Measured review time with the tool against the current manual process.
  7. A re-test plan for every software update, stain change or new instrument.

FDA had authorized 1,451 AI-enabled medical devices by the end of 2025, according to MedTech Dive. Clearance tells you the vendor met a regulatory bar. It does not tell you how the tool performs on your patients, your stains and your case mix. Local verification does.

How Acumen helps

Acumen builds hematology AI evaluations with credentialed hematologists and hematopathologists matched to the task. We assemble enriched test sets, label cells and cases with independent multi-reviewer adjudication, write case-specific rubrics for decision support and grade model outputs against them. Every judgment carries its reviewer record, so the evidence holds up in validation files and procurement review.

See how our medical model evaluation works, then start with a pilot on the cell classes or case types you plan to launch first.

Request a sample dataset

For related reading, see our guides to running a physician review study, evaluating AI skin cancer detection and evaluating AI mammography.

FAQ

What is AI in hematology?

AI in hematology covers software that analyzes blood smears and bone marrow aspirates, pre-classifies cells and estimates counts, plus newer tools that combine smear, CBC and flow data into decision support. In clinical labs today, the AI drafts cell calls and a technologist or pathologist confirms them before results are released.

What is digital morphology in hematology?

Digital morphology is the scanning of stained blood or marrow slides into images that software locates and pre-classifies cell by cell. A trained reviewer then verifies or corrects each call on screen. It replaces much of manual microscope counting but still depends on expert verification, especially for abnormal and rare cells.

How accurate is AI blood smear analysis?

Accuracy varies by tool and cell class. A 2025 Clinical Chemistry and Laboratory Medicine study of 461 slides found 90.0% overall accuracy for one analyzer and 45.5% for another. Both performed worse with abnormal cells, and blast and atypical lymphocyte errors were common, so per-class results matter more than overall accuracy.

Can AI detect blasts on a blood smear?

AI can flag possible blasts, but it should not be the final word. In a 2025 comparison, one analyzer over-called blasts at a 30.5% false positive rate and another missed 42.1%. Guidance from hematology standards bodies, cited by the study authors, recommends manual review whenever blasts are observed.

Who should label ground truth for hematology AI?

Experienced hematologists, hematopathologists or senior morphology technologists should label it, with at least two independent reviewers per cell or case and a third expert adjudicating disagreements. Measure their agreement first. Without multi-reviewer consensus, the AI is graded against one person's habits rather than an expert standard.

How should a lab validate a hematology AI tool before go-live?

Agree acceptance criteria in advance, then test on an independent, enriched slide set from your own stainers and sites. Report per-class sensitivity and predictive value, blast sensitivity, flag rates and review time. Route every blast call to manual review, and re-test after software updates, stain changes or new instruments.